Aptamer-based Microfluidic Device for Isolation of Circulating Tumor Cells
Автор:
Lukyanenko, Kirill A.
Koshmanova, Anastasia A.
Shesternya, Pavel A.
Kichkailo, Anna S.
Лукьяненко, Кирилл А.
Кошманова, Анастасия А.
Шестерня, Павел А.
Кичкайло, Анна С.
Дата:
2024-12Журнал:
Журнал Сибирского федерального университета. Математика и физика 2024 17(6). Journal of Siberian Federal University. Mathematics & Physics 2024 17(6)Аннотация:
The paper presents the design of a microfluidic device for capturing circulating tumor cells,
a method for immobilizing aptamers on the surface of the chip reaction chamber, and an algorithm for
controlling flows in microchannels. It was shown that the maximum efficiency of aptamer immobilization
was achieved by treating the chip surface with a 50% aqueous-alcoholic NaOH solution. In addition, it
was found that the most number of tumor cells of the MCF7 culture attached on the surface coated with
aptamers at low flow rates of less than 2 l/min. It was noted that immobilized tumor cells are capable
of being retained by aptamers at flow rates of up to 200 l/min В работе представлен дизайн микрофлюидного устройства для улавливания циркулирующих опухолевых клеток, способ иммобилизации аптамеров на поверхности реакционной
камеры чипа и алгоритм управления потоками в микроканалах. Показано, что максимальная эффективность иммобилизации аптамеров достигается при обработке поверхности чипа 50% водно-спиртовым раствором NaOH. Кроме того, установлено что наибольшее количество опухолевых
клеток культуры MCF7 осаждается на поверхности, покрытой аптамерами, при небольших скоростях потока менее 2 мкл/мин. Отмечено, что иммобилизованные опухолевые клетки способны
удерживаться аптамерами при скоростях потока вплоть до 200 мкл/мин